Hanna Instruments hi 9811 사용자 설명서

다운로드
페이지 16
9
For temperature compensation during calibration, please refer to the
following chart.
pH VALUES AT VARIOUS TEMPERATURES
For instance, if the buffer temperature is 25°C, the display should
show pH 4.0 or 7.0 or 10.0.
If the buffer temperature is 10°C, the display should show pH 4.0 or
7.0 or 10.1.
Note: The probe should be submerged approximately 4 cm (1½") into the
solution. The ChecktempC thermometer has to be located close to
the  probe.
Note: If turning the knob the needed value can not be reached,
clean the probe (see the "Probe Maintenance" section).
If also after the probe cleaning the value can not be reached,
replace the probe.
TEMP
pH VALUES
°C
°F
4.01
6.86
7.01
9.18
10.01
0
5
10
15
20
25
30
35
40
45
50
55
60
65
70
32
41
50
59
68
77
86
95
104
113
122
131
140
149
158
4.01
4.00
4.00
4.00
4.00
4.01
4.02
4.03
4.04
4.05
4.06
4.07
4.09
4.11
4.12
6.98
6.95
6.92
6.90
6.88
6.86
6.85
6.84
6.84
6.83
6.83
6.84
6.84
6.85
6.85
7.13
7.10
7.07
7.04
7.03
7.01
7.00
6.99
6.98
6.98
6.98
6.98
6.98
6.99
6.99
9.46
9.39
9.33
9.27
9.22
9.18
9.14
9.10
9.07
9.04
9.01
8.99
8.97
8.95
8.93
10.32
10.24
10.18
10.12
10.06
10.01
9.96
9.92
9.88
9.85
9.82
9.79
9.77
9.76
9.75