Hanna Instruments hi 9835 사용자 설명서

다운로드
페이지 11
7
6
EC Calibration
1 point with 6 memorized buffers
84,1413,5000,12880,80000,111800 µS/cm
NaCl Calibration
1 point with 
HI 7037 buffer (optional)
Temperature
2 points at 0 and 50 C
Calibration
(plus ±1 C adjustment)
Temperature
Automatic or Manual from 0 to 60 C
Compensation
(can be disabled to measure actual conductivity)
Temperature
0.00 to 6.00 %/ C (for EC and TDS only)
Coefficient
Default value is 1.90%/ C
TDS Factor
0.40 to 0.80 (default value is 0.50)
Probe
HI  76309 4-ring probe, K=1 nominal and
built-in temperature sensor (included)
Auto Off
After 5 minutes (can be disabled)
Power supply
4x1.5V AA alkaline type batteries (included), or
12 VDC adapter
Casing
IP 67
Environment
0 to 50 C ; 100% RH
Dimensions
196 x 80 x 60 mm (7.7 x 3.1 x 2.4")
Weight
500 g (18 oz)
CONNECTIONS
CONNECTIONS
CONNECTIONS
CONNECTIONS
CONNECTIONS
Remove the back cover, unwrap the batteries
and install them while paying attention to
the polarity. Replace the back cover.
Alternatively, connect the voltage adapter to
the power socket.
Connect the EC/TDS probe to the
7-pin connector located on the top
of the instrument. Tighten the
threaded ring.
Make sure the probe sleeve is prop-
erly inserted, as shown in the figure.
Press the ON/OFF key to turn the meter on.
Immerse the probe into the solution to be tested.
The sleeve holes must be completely submerged.
Tap the probe repeatedly to remove any air bubbles
that may be trapped inside the sleeve.
If needed, press the RANGE key repeatedly until
the desired range (EC, TDS, NaCl) is selected on
the LCD.
Allow for the reading to stabilize. The upper
LCD displays the measure in the selected
range while the temperature is displayed on
the lower LCD.
Notes:
I
If the meter displays "
----" the reading is out of range.
II
If the reading is unstable,  the stability indicator " " blinks.
III
The "gm" indication on the LCD means g/L.
IV
Make sure the meter is calibrated before taking measurements.
V
If measurements are taken successively in different samples, to
have accurate readings it is recommended to rinse the probe
thoroughly with deionized water before immersion in the samples.
VI
To maximize battery life, the meter is automatically switched off
after 5 minutes of non-use. To reactivate the instrument press
the ON/OFF key. This feature can be disabled by entering the
setup mode and selecting the "AoF" item (see SETUP for de-
tails).
VII TDS reading is obtained multiplying the EC reading by the TDS
factor, which has a default value of 0.50. It is possible to
change the TDS factor in the 0.40 to 0.80 range by entering
the setup mode and selecting the "tdS" item (see SETUP for
details).
VIII When the use of an alternate function (FNC, CFM, GLP,FIXED,TC
and CALT) is requested, press and hold the ALT key first and
then the second key.
TAKING MEASUREMENTS
TAKING MEASUREMENTS
TAKING MEASUREMENTS
TAKING MEASUREMENTS
TAKING MEASUREMENTS