Cadmium Electrode 
 
Instruction Manual
 
 
12
Temperature Influences 
 
Samples and standards should be at the same temperature, since the electrode potentials are 
influenced by changes in temperature. A 1
o
C difference in temperature results in a 4% error at the 
1.0X10
-3
M cadmium concentration. Because of the solubility equilibrium on which the electrode 
depends, the absolute potential of the reference electrode changes slowly with temperature. The 
slope of the electrode, as indicated by the factor "S" in the Nernst equation, also varies with 
temperature. Table 3 gives values for the "S" factor in the Nernst equation for the cadmium ion. 
 
TABLE 3:  Temperature vs. Values for the Electrode Slope 
 
                    Temp.(oC)            
  "S" 
                        0 
 
 
27.10 
                       10 
 
 
28.10 
                       20 
 
 
29.08 
                       25 
 
 
29.58 
                       30 
 
 
30.07 
                       40 
 
 
31.07 
                       50 
 
 
32.06 
 
If changes in temperature occur, the electrodes should be re-calibrated. The temperature range for 
the Eutech Cadmium Ion Electrode is 0
o
-80
o
C, provided that temperature equilibrium has occurred. 
If the temperature varies substantially from room temperature, equilibrium times up to one hour are 
recommended. 
 
Electrode Response 
 
Plotting the mV potential against the cadmium concentration on semi-logarithmic graph paper 
results in a straight line with a slope of about 27 mV per decade. (Refer to Figure 1.) 
 
The time needed to reach 99% of the stable electrode potential reading, the electrode response time, 
varies from several seconds in highly concentrated solutions to several minutes near concentrations 
of 1.0X10
-6
M cadmium. Below 10
-6
M, considerably longer response time can be expected.  (Refer 
to Figure 3.)