Instruction Manual
EL 9000 HP Series
Date: 10-31-2011
6.9  Control locations and priorities
Control locations are the locations from where the device 
is controlled. This can be at the device (manual control), 
via the analogue interface (external control) or via a di-
gital interface card (remote control). In order to prevent 
the user from accessing the device from two locations at 
once there are priorities. Following applies:
The analogue interface has the highest priority, the digital 
interface the second highest one and the manual control 
the lowest one. It means, that if the device was set to re-
mote control, modes and set values can‘t be set with the 
switches and rotary knobs anymore. If the device would 
be switched to external control while remote control is 
active, the remote control status would be reset and the 
device would only be controllable via the analogue inputs. 
In order to report this to a software running on a PC, still 
trying to access and control the device, the control loca-
tion is internally set to „local“. In „local“ status the device 
can only be read (i.e. monitored) by the PC.
6.10  Series and parallel connection
Parallel connection of multiple loads is possible, but it is 
not explicitely supported. That means, there is no auto-
matic distribution of the current at parallel connection. The 
user has to take care of the correct control of the devices.
When using parallel connection, the symmectric distri-
bution is achieved by adjusting the same set values for 
U,I,P and R at any device via the control panel or the 
interfaces (digital or analogue).
Attention!  Series  connection  is  not  allowed!  The 
devices could be damaged.
Handling the device 
Figure 13. Battery test operation in current control (CC) mode
6.11  Functions of the connector „System Bus“
6.11.1 Two-quadrants operation
The Share Bus of the electronic load is compatible for 
so-called  two  quadrants  operation  with  certain  power 
supply series:
•  Series PS 9000
•  Series PSI 9000
•  PSI 8000 DT/2U
•  PS 8000 DT/2U
Other series or subseries, even if they do feature a Share 
bus, are not compatible or only by extra measures.
The connection for two quadrants operation is as follows:
PS 9000 / PSI 9000:
Connect pin 5 (Share Bus) of the EL9000 to pin 5 (Share 
Bus) of the power supply and pin 6 (AGND) of the EL9000 
with pin 6 (AGND) of the power supply.
PS 8000 / PSI 8000:
Connect pin 5 (Share Bus) of the EL9000 to pin 7 (Share 
Bus  +)  of  the  power  supply  and  pin  6  (AGND)  of  the 
EL9000 with pin 6 (AGND or Share Bus -) of the power 
supply.
In  Share  Bus  operation,  the  electronic  load  operates 
as the leading component and the power supply as the 
depending one.
Typical applications for the two-quadrants operation are:
•  Battery tests with semi-automatic charge and discharge 
cycles
•  Automotive electronic tests with simulation of transients 
like for example voltage break-ins during engine starts
•  Cyclic charging and discharging of capacitors
Note:  two-quadrants  operation  is  only  suitable  for  CV 
mode (constant voltage).