Annex page 3 of 7 
 
Annex to Test Certificate No. DK0199-R76-11.10 
Issued by DELTA 
11), and any particular installation requirements. A load cell marked NH is allowed only if 
humidity testing to EN 45501 has been performed. 
• 
The compatibility of load cells and indicator is established by the manufacturer by means of 
the compatibility of modules form, contained in the above WELMEC 2 document, or the like, 
at the time of EC verification or declaration of EC conformity of type. 
• 
The load transmission must conform to one of the examples shown in the WELMEC 2.4 Guide 
for load cells. 
4.2 
Peripheral interfaces 
The indicator may be equipped with the following protective interfaces that have not to be secured. 
• 
RS-232C 
The peripheral interfaces are characterised “Protective interfaces” according to paragraph 8.4 in the 
Directive. 
5. 
Conditions for use 
Legal use of the indicator for automatic weighing or as counting device is not allowed with reference 
to this test certificate. 
6. 
Location of seals and inscriptions 
Seals shall bear the verification mark of a notified body or alternative mark of the manufacturer ac-
cording to ANNEX II, section 2.3 of the Directive 2009/23/EC. The seals shall be placed as shown in 
section 9. 
Location of CE mark of conformity: 
The CE mark of conformity is placed on the overlay on the rear side of the device. 
Inscription near the display: 
Max
i
, Min
i
, e
i
  
Inscription on the overlay: 
Type, accuracy class, Temp. -10 °C / +40 °C, Certificate No. DK0199-R76-11.05. 
Other inscriptions on the overlay: 
Manufacturer’s name and/or logo, Part No, Supply voltage.