Sentry Industries 5000 사용자 설명서

다운로드
페이지 130
SMC
 sierra monitor corporation 
Sentry Instruction Manual - Version 6 
 
CONFIGURATION PROCEDURE 
Page:  41 
SENTRY KEY BOARD 
Ts = TEST  R = RESET  M = MODE  T = TIME 
^ = ARROWS  E = ENTER 
 
 
 
 
 
 
 
RESET KEY 
 
 
*  GLOBAL CALIBRATION 
 
 
Reset Alarms 
R  E 
 
 
Gas Type (Concentration) 
 
 
Reset System 
R  R  E 
 
 
Enter to Span 
 
 
Reset History 
R  R  R  E 
 
 
Apply Zero Gas 
 
 
TIME KEY 
 
 
Enter to Span 
 
 
Check History 
T  E 
 
 
Apply Span Gas 
 
 
 
Printer History 
T  T  E 
 
 
Operate Mode (After timer) 
 
 
 
Print System 
T  T  T  E 
 
 
*  LOCAL CALIBRATION 
 
 
 
Print Status 
T  T  T  T  E 
 
 
Gas Type (Concentration) 
^ E 
 
 
Set Date & Time 
T  T  T  T  T  E 
 
 
Apply Zero & Span 
 
 
TEST KEY 
 
 
Operate Mode (After timer) 
 
 
Test System 
Ts  E 
 
 
CHANGE MODULE 
 
 
Test Alarms 
Ts  Ts  E 
 
 
Module # 
 
 
MODE KEY - PASSIVE 
 
 
Gas Type 
 
 
Change/Calib 
M  E 
 
 
Module On/Off 
 
 
Select Scan 
M  M  E 
 
 
Low Alarm Latch/Non 
 
 
Check Calib 
M  M  M  E 
 
 
High Alarm Latch/Non 
 
 
Check Module 
M  M  M  M  E 
 
 
High Alarm Value 
E  ^ 
E  ^ 
 
 
MODE KEY - ACTIVE 
 
 
Low Alarm Value 
E  ^ 
E  ^ 
 
 
Change/Calib 
M  E 
 
 
Scaling Factors 
E  ^ 
E  ^ 
 
 
Calibrate 
* 
 
 
*  SET USER CODE 
 
 
Change Calibrate 
* 
M  E 
 
 
User Number 
 
 
Change Module 
* 
M  M  E 
 
 
Entry Code 
^E  ^E  ^E  ^E 
 
 
Set User Code 
* 
M  M  M  E 
 
 
*  CHANGE PRINTER 
 
 
Change Printer 
* 
M  M  M  M  E 
 
 
Printer On/Off 
 
 
Change Gas Tag 
 
M  M  M  M  M  E 
 
 
Controller ID 
^^  E 
 
 
Change Module Tag 
 
M  M  M  M  M  M  E 
 
 
Status Frequency 
^^  E 
 
 
*  CHANGE CALIB  
 
 
Print History 
 
 
Gas Type 
^^  E 
 
 
Gas Type 
^^  E 
 
 
Full Scale 
^^  E  ^^  E  ^^  E 
 
 
Min. Conc. 
^^  E 
 
 
Calib Concentration 
^^  E  ^^  E  ^^  E 
 
 
Conc. Change 
^^  E 
 
 
Calib Frequency 
^^  E 
 
 
Gas Type 
 
 
Global Calib 
 
 
Exit 
 
Figure 4-2 
Sentry Keyboard - Quick Help