7.3.6  DIAGNOSTIC REPORT 
A Diagnostic report can be printed using diagnostic codes 
0008 or 0010. Figure 7-4 is a typical report.. 
ID:A 11/20 15:16 
S1= +0.376 
BV= +2.021 
BI= 0.323 
SP= 0653 
 
ST= +0.284 
MV= +19.0 
CS= +0.000 
S2= +1.20 
HV= +0.920 
HI= +0.975 
SP= 400 
 
ST= +0.009 
MV= +19.4 
CS= +0.000 
Figure 7-4 
Typical Diagnostic Report 
Table 7-2 is a description of the information contained in 
a typical diagnostic report 
Report 
Description 
 
ID:A 
Controller Identification 
 
11/20 15:16 
Time report was generated 
 
S1 
+0.376 
Data that follows is for Module 1 
Sensor Signal is +0.376 VDC 
 
BV 
+2.021 
Bridge Voltage 2.021 VDC 
 
BI 
+0.322 
Bridge Current 323 mA 
 
SP 
  0653 
Sensor Power 653 mW 
 
ST 
+0.284 
Sensor Type 
(Where: .2 -.3 VDC: Combustible) 
 
MV 
+19.0 
Module Voltage = 19 VDC 
 
CS 
+0.000 
Calibration Switch 
Off 0.000, On = 2.5 VDC 
 
S2 
+1.20 
Data that follows is for Module  
Sensor Signal is + 1.20 VDC 
 
HV 
+0.920 
Heater Voltage Signal 
 
HI 
+0.975 
Heater Current Signal 
 
SP 
  400 
Sensor Power 400 mW 
 
ST 
+0.009 
Module Type 
(Where: .11-.1 VDC: H2S) 
Table 7-2 
Diagnostic Report Description