Proline Promass 80 PROFIBUS PA
Troubleshooting
79
9.4
Process errors without messages
Symptoms
Rectification
!
Note! 
You may have to change or correct certain settings of the function matrix in order to rectify faults. The functions outlined below are described in detail in the 
"Description of Device Functions" manual.
Measured value reading fluctuates even 
though flow is steady.
1.
Check the fluid for presence of gas bubbles.
2.
Increase the following values:
– Analog Input function block 
→ RISING TIME
– BASIC FUNCTIONS 
→ SYSTEM PARAMETER → CONFIGURATION → FLOW DAMPING
3.
Increase the value for display damping:
HOME 
→ USER INTERFACE → CONTROL → BASIC CONFIG. → DISPLAY DAMPING
Measured value reading shown on 
display, even though the fluid is at a 
standstill and the measuring tube is full.
1.
Check the fluid for presence of gas bubbles.
2.
Enter a value for the low flow cutoff or increase this value:
BASIC FUNCTION 
→ PROCESS PARAMETER → CONFIGURATION → ON-VALUE LF CUTOFF
The fault cannot be rectified or some 
other fault not described above has 
occurred.
In these instances, please contact your 
Endress+Hauser service organization.
The following options are available for tackling problems of this nature:
Request the services of an Endress+Hauser service technician
If you contact our service organization to have a service technician sent out, please be ready with the following 
information:
• Brief description of the fault
• Nameplate specifications: order code and serial number 
Return devices to Endress+Hauser
You can return a measuring device to Endress+Hauser for repair or calibration. 
Always enclose the duly completed "Declaration of contamination" form with the flowmeter. You will find a preprinted 
blank of this form at the back of this manual.
Replace transmitter electronics
Parts of the measuring electronics defective 
→ Order spare parts → Page 80 ff.