Philips MP40/50 Manual Do Utilizador

Página de 414
Perfusion (Pleth) Indicator
9 Monitoring  SpO
2
141
 
Perfusion (Pleth) Indicator
The perfusion indicator gives a numerical value for the pulsatile portion of the measured signal caused 
by the pulsating arterial blood flow.
As pulse oximetry is based on the pulsatile nature of the signal, you can also use the perfusion indicator 
as a quality indicator for the SpO
2
 measurement. Above 1 is optimal, between 0.3-1 is acceptable. 
Below 0.3 is marginal; reposition the sensor or find a better site.
Setting SpO
2
/Pleth as Pulse Source
1
In the 
Setup SpO2
 menu, select 
Pulse (SpO
2
to enter the 
Setup Pulse 
menu.
2
In the 
Setup Pulse 
menu, select
 System Pulse 
and select 
SpO
2
 from the pop-up list.
Setting Up Tone Modulation
If tone modulation is on, the QRS tone pitch lowers when the SpO
2
 level drops. Remember, the QRS 
tone is derived from either heart rate or pulse depending on which is currently selected as the active 
alarm source. 
In the 
Setup
 
SpO
2
 
menu
select
 Tone Modulation
 to toggle between 
Yes
 (for on) and 
No
 
(for off ).
Tone modulation is licensed under US patent US 4,653,498 from Nellcor Puritan Bennett 
Incorporated, a Tyco Healthcare company.
Setting the QRS Volume
In the 
Setup SpO
2
 
menu,
 
select
 QRS Volume 
and set the appropriate QRS tone volume.
Calculating SpO
2
 Difference
When a second SpO
2
 measurement is present (either through the SpO
2
 module or through VueLink), 
the monitor displays both SpO
2
 values, and calculates the difference between them. The second value 
is subtracted from the first.
minimum size for reliable SpO
2
 value