ResMed 338425/1 Manual Do Utilizador
32
System Specifications
Noise Level
Accessories
Data Card
Dimensions (L x W x D): 65.04 mm x 22.15 mm x 0.85 mm
Data Card Module
Dimensions (H x W x D): 49.7 mm x 42.2 mm x 12 mm
All Devices (where applicable)
Housing Construction: Flame retardant engineering thermoplastic
Environmental Conditions
Operating Temperature: +5°C to +36°C
Operating Humidity: 10%–95% non-condensing
Storage and Transport Temperature: -20°C to +60°C
Storage and Transport Humidity: 10%–95% non-condensing
Operating Altitude: sea level to 2591 m
Operating Humidity: 10%–95% non-condensing
Storage and Transport Temperature: -20°C to +60°C
Storage and Transport Humidity: 10%–95% non-condensing
Operating Altitude: sea level to 2591 m
Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments. For further details, see “Guidance and Manufacturer’s
Declaration — Electromagnetic Emissions and Immunity” on page 34.
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments. For further details, see “Guidance and Manufacturer’s
Declaration — Electromagnetic Emissions and Immunity” on page 34.
Note: The manufacturer reserves the right to change these specifications
without notice.
without notice.
AutoSet Spirit II—CPAP mode, 10 cm H
2
O set pressure
*
*
Declared dual number noise emission values in accordance with ISO 4871.
Measured total A-weighted sound pressure level
(dB re 20 μPa)
(dB re 20 μPa)
24 dBA
Uncertainty (dB re 20 μPa)
2 dBA
Measured values are determined according to the noise test procedure given in
ISO 17510-1.
ISO 17510-1.
Note: The sum of a measured noise emission value and its associated
uncertainty represents an upper boundary of the range of values that is
likely to occur in measurements.
uncertainty represents an upper boundary of the range of values that is
likely to occur in measurements.