Honeywell 96D Manual Do Utilizador

Página de 80
M-511401
  96d User Manual 
49
8/10
Honeywell 
System Programming
Acquisition Menu
Starting and Stopping Tx Logging
In the previous step, “Acquisition”, the frequency at which Tx logs would 
be recorded can be configured. To start the logging function:
When “Start Tx logging” appears on the display, it  indicates that the 
acquisition, or logging, mode is inactive.  When “Stop Tx logging” 
appears, it indicates that Tx data is being logged.  The log message is 
displayed on the screen according to the chosen mode and LED 1 will 
light up.
Press the Enter keypad button to stop or start Tx logging.
When Tx data is logged, the system creates files named 
tayymmdd.log, tbyymmdd.log and tcyymmdd.log, each representing 
one third of the network.  The record includes the transmitter’s date, 
time and address, the sensor type, the concentration read, as well as 
the alarm status.  Here is a sample of what a Tx log looks like:
The first column of the Tx log displays the date (yyyy-mm-dd) and the 
time (hh:mm:ss) of the log.  In this example, the “Delay mode” was set 
to one minute intervals.
The third column of the Tx log displays the transmitter address and the 
fourth displays the gas type, gas concentration and unit of 
measurement.  
The display then lists the next transmitter address with its gas type, 
concentration and unit of measurement, and so on until all the 
transmitters have been listed.
2004-01-23 17;54;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;1.5;-normal-;003_CO_ppm;0;-normal-
2004-01-23 17;55;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;0.5;-normal-;003_CO_ppm;0;-normal-
2004-01-23 17;56;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;0.5;-normal-;003_CO_ppm;0;-normal-
2004-01-23 17;57;25; 001_CO_ppm;0;-normal-;002_NO2_ppm;1.0;-normal-;003_CO_ppm;0;-normal-
2004-01-23 17;58;25;  001_CO_ppm;0;-normal-;002_NO2_ppm;1.5;-normal-;003_CO_ppm;0;-normal-