Teledyne 201A Manual Do Utilizador

Página de 31
Teledyne API Model 201A NH
3
 Analyzer Instruction Manual, 04689, Rev. B1 
 
1.2 M201A Analyzer Specifications 
Ranges 
0-50 to 0-2000 ppb in 1 ppb increments 
Range Modes 
Single, Independent, AutoRange 
Noise at Zero
1
0.5 ppb RMS 
Noise at Span
1
< 1.0% of reading above 50 ppb 
Lower Detectable Limit  1 ppb RMS 
Zero Drift
3
2 ppb / 24 hours 
Span Drift 
< 1.0% FS  Range / 7 days  
Lag Time 
40 sec 
Rise Time 
90% 120 sec 
Fall Time 
90% 120 sec 
Sample Flow Rate 
1000 cc/min 
Linearity 
2% of full scale 
Precision 
1.0% of reading 
Temp Range 
10-30
o
Dimensions HxWxD 
7” x 17” x 23.6” (18cm x 43cm x 61cm) 
Weight, Analyzer 
43 lbs (20 kg) 
Weight, Converter 
24 lbs (11 kg) 
Weight, Pump 
16 lbs (7 kg) 
Power, Analyzer 
100V ~50/60 Hz, 120V ~60 Hz, 220V ~50Hz, 240V ~50 Hz, 
125 watts 
Power, Analyzer
5
230V ~50 Hz, 125 watts 
Power, Pump 
110V ~60 Hz, 220V ~50 Hz, 240V ~50 Hz, 295 watts 
Power, Pump CEMark
5
230 V ~50 Hz, 2.5 A peak 
Environmental Installation 
Category Pollution Degree 2, Over-voltage 
Category II 
Analog Output 
0-100, 0-1, 0-5, 0-10VDC bi-polar, 4-20mA isolated 
Analog Resolution  
1 part in 2048 
Status Option 
12 Status Outputs 
Measurement Units 
ppb, ug/m3 
 
1. As defined by USEPA 
2. Defined as twice the zero noise level 
3. At constant temperature and voltage 
4. Bi-polar 
5. Electrical rating for CE Mark Compliance