Label Gone Timeout
Product Reference Guide
         25
Label Gone Timeout
This feature sets the time after the last label segment is seen before the reader prepares for a new 
label. The timeout can be set within a range of 10 milliseconds to 2,550 milliseconds (2.55 
seconds) in 10ms increments. Label Gone Timeout does not apply to scan modes that require a 
trigger pull for each label that is read
Follow these instructions to set this feature:
1. Determine the desired setting in milliseconds.
2. Divide the desired setting by 10 (setting is in 10ms increments). Pad the result with lead-
ing zeroes to yield three digits. For example: 0 = 000, 5 = 005, 20 = 020, etc.
3. Scan the ENTER/EXIT PROGRAMMING MODE barcode to enter Programming 
Mode.
4. Scan the barcode: SELECT LABEL GONE TIMEOUT SETTING.
5. Scan the appropriate three alpha-numeric characters from the keypad in 
Appendix E, Key-
pad
 representing the duration which was determined in the steps above. You will hear a 
two-beep indication after the last character.
NOTE
6. Scan the ENTER/EXIT PROGRAMMING MODE barcode to exit Programming 
Mode.
This completes the procedure. See Table
2
 for some examples of how to set this feature.
Table 2. Timeout Setting Examples
STEP
ACTION
EXAMPLES
1
Desired Setting
50ms
150ms
1800ms (1.8 sec.)
2550ms (2.55 sec.)
2
Divide by 10 (and pad with 
leading zeroes)
005
015
180
255
3
Scan ENTER/EXIT PROGRAMMING MODE
4
Scan SELECT LABEL GONE TIMEOUT SETTING
5
Scan Three Characters 
From 
Appendix E, Keypad
‘0’, ‘0’ and ‘5’
‘0’, ‘1’ and ‘5’
‘1’, ‘8’ and ‘0’
‘‘2’, ‘5’ and ‘5’
6
Scan ENTER/EXIT PROGRAMMING MODE
If you make a mistake before the last character, scan the CANCEL 
barcode to abort and not save the entry string. You can then start 
again at the beginning.