Honeywell SC1000 Manual Do Utilizador

Página de 174
 page 136      008-0608-00
14.7  Troubleshooting
14.7.1 Error Messages
See “Error Messages” on page 151 for information relating to error messages.
14.7.2 Common 
Problems and Solutions
Analog Output Incorrect
Make certain of the type of Analog Output (voltage or current) that the channel is
equipped with; see “Identifying the Output Type” on page 135.
Use the 
DIAGNOSTICS 
-> 
DAC FULL SCALE
 and 
DIAGNOSTICS
 -> 
DAC ZERO SCALE
 menu items to
force the Analog Output to a known output.  Then, adjust your readout device, panel
meter, PLC or data acquisition system to match.
Check the output selection jumpers; see “Output Selection” on page 135.
Check the “
DAC SETUP
 -> 
DAC. ZERO-SCALE
” and “
DAC SETUP
 -> 
DAC FULL-SCALE
” menu items.
Sensitivity to EMI/RFI
To obtain maximum immunity to electromagnetic or radio frequency interference,
make certain that the shield of the cable is connected to ground/earth only at the read-