Honeywell SC1000 Manual Do Utilizador
page 80 008-0608-00
11.2 Wiring
Connect your transducer to an AC/AC-LVDT Input channel by wiring it to the 12-pin
connector of that channel. The Customer Information Sheet that shipped with the
instrument describes which cards are installed in each channel. The pin-out for this
connector is shown on the following table.
connector of that channel. The Customer Information Sheet that shipped with the
instrument describes which cards are installed in each channel. The pin-out for this
connector is shown on the following table.
The Analog Output and Analog Return pins are electrically isolated from all other pins
on the instrument.
on the instrument.
Figure 11-1: Connection of Four- or Five-wire AC/AC-LVDT
Table 5-5: AC/AC-LVDT Input Channel Pin Connections
Pin
Label
Function
Reference Pin
1 (top)
+EXC
(+)Excitation
10
2
N/C
No connection
3
N/C
No connection
4
-EXC
(-)Excitation
10
5
+SIG
(+)Signal
10
6
-SIG
(-)Signal
10
7
+OUT
Analog Output
8
8
-OUT
Analog Return
-
9
N/C
No connection
10
DGND
Digital Ground
-
11
AUX1
Auxiliary Function 1
(connect to pin 10 to activate)
10
12
(bottom)
AUX2
Auxiliary Function 2
(connect to pin 10 to activate)
10