SC Series Instruction Manual     page 89
AC
/A
C-L
VDT I
nput Ch
annel
11
11.5.6 
CALIBRATE
 Menu 
Item
This menu item performs a calibration according to what was entered in the 
CALI-
BRATION TYPE
 and 
CALIBRATION DATA
 menu items.
If the 
CALIBRATION TYPE
 is 2-Point Known Displacement Calibration...
The display will read 
DOING 2POINT CAL
, and prompt you to 
APPLY 00000 UNIT
 
(where 00000 and UNIT are previously entered as
 KNOWN POINT 1/2
 and 
DISPLAY. 
UNITS
). When you have applied this displacement to the transducer, press 
[ENTER]. 
The display will read 
WORKING
, then 
APPLY 01000 UNIT
 (where 01000 and UNIT are 
previously entered as
 KNOWN POINT 2/2
 and 
DISPLAY. UNITS
). When you have 
applied this displacement to the transducer, press [ENTER].
Upon completion, the display will indicate 
DONE
 and the instrument will return to 
the RUN mode.
If the 
CALIBRATION TYPE
 is 3-Point Known Displacement Calibration...
The display will read 
DOING 3POINT CAL
, and prompt you to 
APPLY 00000 UNIT
 
(where 00000 and UNIT are previously entered as
 KNOWN POINT 1/3
 and 
DISPLAY. 
UNITS
). When you have applied this displacement to the transducer, press 
[ENTER]. 
The display will read 
WORKING
, then 
APPLY 00500 UNIT
 (where 00500 and UNIT are 
previously entered as
 KNOWN POINT 2/3
 and 
DISPLAY. UNITS
). When you have 
applied this displacement to the transducer, press [ENTER]. 
The display will read 
WORKING
, then 
APPLY 01000 UNIT
 (where 01000 and UNIT are 
previously entered as
 KNOWN POINT 2/3
 and 
DISPLAY. UNITS
). When you have 
applied this displacement to the transducer, press [ENTER]. 
Upon completion, the display will indicate 
DONE
 and the instrument will return to 
the RUN mode.
If the 
CALIBRATION TYPE
 is 5-Point Known Displacement Calibration...
The display will read 
DOING 5POINT CAL
, and prompt you to 
APPLY 00000 UNIT
 
(where 00000 and UNIT are previously entered as
 KNOWN POINT 1/5
 and 
DISPLAY. 
UNITS
).  When you have applied this displacement to the transducer, press 
[ENTER]. 
The display will read 
WORKING
, then 
APPLY 00250 UNIT
 (where 00250 and UNIT are 
previously entered as
 KNOWN POINT 2/5
 and 
DISPLAY. UNITS
). When you have 
applied this displacement to the transducer, press [ENTER].
The display will read 
WORKING
, then 
APPLY 00500 UNIT
 (where 00500 and UNIT are 
NOTICE
Before performing a calibration,
 the transducer must be connected to the instrument,  the 
CALIBRATION TYPE
 must 
CALIBRA-
TION DATA
NOTICE
For maximum accuracy, allow at least twenty minutes of warm-up with the exci-
tation voltage applied to the transducer before calibration.