8.8.2  P7.2: Setting the normalisation temperature 
Your meter will normalise its conductivity measurements to a 
standard temperature that you can select. You can adjust 
the normalisation temperature either at 20.0°C or 25.0°C. 
Meter default is 25.0°C.  
From measurement mode 
1.  Press SETUP key to enter Set Up menu. 
2.  Press the MI/▲ or MR/▼ keys until you view 
parameter P7.0.  
3.  Press the HOLD/ENTER key three times to select 
display “tnr” at parameter P7.2.  
4.  Press the HOLD/ENTER key again. The upper 
display shows the normalisation temperature of 25.0 
°C. 
5.  Press the MI/▲ or MR/▼ keys to select either 
20.0°C or 25.0°C as the normalisation temperature.  
6.  Press the HOLD/ENTER key to confirm selection 
and to return to the subgroup menu.  
7.  Press CAL/MEAS key to return to measurement 
mode.  