28
8.2 
Program 1: Common functions
The following 3 options are commonly used:
8.2.1 
P1.0: Memory Clear
16 sets of readings can be stored in the meter. Both
conductivity and temperature or TDS and temperature
readings are stored.
ON
Clears all values stored in memory
OFF
Memory values remain in storage
Activation of this option by selecting ON and confirmation of
memory clears all memory values. This is done by toggling
the annunciator to ON using the MI/
!
 or MR/
"
 key, and
pressing ENTER/RANGE to confirm the selection.
8.2.2 
P1.1: %Temperature coefficient
adjustment
The amount of change in conductivity per degree centigrade
is referred to as the Temperature Coefficient, expressed in
%/
°
C at a particular temperature. The meter provides the
options of selecting the appropriate temperature coefficient,
from 0.0 to 10.0 %/
°
C, depending on the type of solution
measured.
The primary display indicates the temperature coefficient in
%/
°
C. You can use the MI/
!
 or MR/
"
 key to scroll to a
required value from 0.0 to 10.0  %/
°
C. Once you have made
your selection, press ENTER/RANGE key to confirm the
value.
The program will automatically scroll to the next SETUP
parameter. Under default condition, the temperature
coefficient is set to 2.0%/
°
C.
Figure 23: Memory clear display
Figure 24: Temperature
coefficient adjustment setup