eutech-instruments calcium epoxy Manual Do Utilizador

Página de 19
Calcium Electrode 
 
Instruction Manual 
 
10
The progressive and quantitative addition of a reagent to a measured sample until neither active 
species (reagent or sample) is in excess. Ion selective electrodes are excellent endpoint detectors 
since they are not influenced by solution color or turbidity. Though titration is more time 
consuming than direct measurement, it is about 10 times more accurate. 
 
Titration of Calcium 
 
The method outlined in this section makes use of the calcium ion electrode as a highly sensitive 
endpoint detector for calcium-containing samples. The titrant used is EDTA. 
 
EDTA complexes calcium as well as other cations. The sample pH can be adjusted to pH 10 by 
adding ammonia to eliminate unwanted ion complexes. Masking agents can be added in some 
cases. 
 
1. 
Soak the calcium ion electrode tip in 10
-3
M or 100 ppm calcium standard solution for a 
minimum of one hour prior to use. 
 
2. 
Prepare the stock EDTA titrant as given in the section 
Required Solutions
. Dilute the 
EDTA to 10 to 20 times as concentrated as the suspected sample concentration. The 
sample should contain at least 1.0X10
-3
M calcium for a good detection of the endpoint. 
 
3. 
Fill a 50 ml  buret with the EDTA solution. Pipet 100 ml  of the sample into a 150 ml  
beaker, place the beaker on the magnetic stirrer and begin stirring at a constant rate. Adjust 
the sample to pH 10 by adding ammonia. 
 
4. 
Position the buret tip in the beaker, slightly above the liquid level in the beaker and slightly 
off center. Position the electrode tips in the solution about half way between the center of 
the beaker and the beaker wall. 
 
5. 
Begin adding the EDTA in 0.5 ml to 1.0 ml increments and about 0.1 ml to 0.2 ml  
increments as the potential begins to change more rapidly. Record the mV potential after 
each addition. Continue the additions several milliliters past the endpoint. 
 
6. 
Plot the milliliters of EDTA added against the mV potential on standard coordinate graph 
paper.  (See Figure 2).  The point of greatest potential change is the endpoint. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
7. 
The calcium ion concentration from the unknown is calculated as follows.