eutech-instruments calcium epoxy Manual Do Utilizador

Página de 19
Instruction Manual 
 
Calcium Electrode 
 
15
Solutions other than this may be used as ISA's as long as ions that they contain do not interfere with 
the electrode's response to calcium ions. 
 
The reference electrode must also be considered. When two solutions of different composition are 
brought into contact with one another, liquid junction potentials arise. Millivolt potentials occur 
from the inter-diffusion of ions into the two solutions. Electrode charge will be carried unequally 
across the solution boundary resulting in a potential difference between the two solutions, since 
ions diffuse at different rates. When making measurements, it is important to remember that this 
potential be the same when the reference is in the standardizing solution as well as in the sample 
solution or the change in liquid junction potential will appear as an error in the measured electrode 
potential. 
 
The composition of the liquid junction filling solution in the reference electrode is most important. 
The speed with which the positive and negative ions in the filling solution diffuse into the sample 
should be equitransferent. No junction potential can result if the rate at which positive and negative 
charge carried into the sample is equal. 
 
 
TROUBLESHOOTING GUIDE 
 
The goal of troubleshooting is the isolation of a problem through checking each of the system 
components in turn: the meter, the glass-ware, the electrodes, the standards & reagents, the sample, 
and the technique. 
 
Meter 
The meter may be checked by following the check-out procedure in the instrument instruction 
manual. 
 
Glass-ware 
Clean glass-ware will drain clean. That is, when rinsed with distilled or deionized water, the water 
does not bead on the inside walls of the glass-ware. 
 
Electrodes 
The electrodes may be checked by using the procedure found in the sections entitled 
Electrode 
Slope Check
 
1. 
Be sure to use distilled or deionized water when following the procedures given in 
Electrode Slope Check.
 
 
2. 
If the electrode fails to respond as expected, see the section 
Measuring Hints
.  Repeat the 
slope check. 
 
3. 
If the electrodes still fail to respond as expected, substitute another calcium electrode that 
is known to be in good working order for the questionable electrode. If the problem 
persists and you are using an electrode pair, try the same routine with a working reference  
electrode. 
 
4. 
If the problem persists, the standards and/or reagents  may be of poor quality, interferences 
in the sample may be present or the technique may be faulty. (See 
Standards and 
ReagentsSample
and
 Technique
 sections below.)