eutech-instruments silversulfide epoxy Manual Do Utilizador

Página de 26
Instruction Manual 
 
Silver/Silver Sulfide Electrode 
 
13
2. 
Dilute 1 ml of 0.1M standard to 1 liter to prepare a 1.0X10
-4
M solution for measurements 
in moles per liter. Prepare a 10 ppm standard solution by diluting 1 ml of the 1,000 ppm 
standard to 100 ml of solution for measurements in ppm. Standards should be prepared 
fresh daily. 
 
3. 
Add 1 ml of low level ISA to a 100 ml volumetric flask and fill to the mark with distilled 
water. Pour this solution into a 150 ml beaker and place the beaker on the magnetic stirrer. 
Begin stirring at a constant rate. 
 
4. 
Place the electrode tips in the solution. Assure that the meter is in the mV mode. 
 
5. 
Add increments of the 1.0X10
-4
M or 10 ppm silver standard as given in Table 2 below. 
 
6. 
After the reading has stabilized, record the mV reading. 
 
     TABLE 2: Stepwise Calibration for Low Level Silver Measurements 
 
 
 
 
 
            Added    
Concentration 
 
   
Step  
Pipet    
Volume (ml)  
ppm 
   M 
    
  A 
 
 
0.1 
 
0.01 
1.0X10-
7
 
 
 
  A 
 
 
0.1 
 
0.02 
2.0X10-
7
 
 
 
  A 
 
 
0.2 
 
0.04 
4.0X10-
7
 
 
 
  A 
 
 
0.2 
 
0.06 
6.0X10-
7
 
 
 
  A 
 
 
0.4 
 
0.10 
9.9X10-
7
  
 
 
 
  B 
 
 
2.0 
 
0.29 
2.9X10-
6
 
 
 
  B 
 
 
2.0 
 
0.48 
4.8X10-
6
 
 
                 Pipet A = 1 ml graduated pipet 
                 Pipet B = 2 ml pipet 
                 Solutions: additions of 10 ppm or 1.0X10-
4
            
        standard to 100 ml of ISA as prepared in Step 3 above. 
 
7. 
On semi-logarithmic graph paper, plot the millivolt reading (linear axis) against the 
concentration (log axis) as in Figure 2. 
 
8. 
Rinse the electrodes in distilled water and blot dry. 
 
9. 
Measure out 100 ml of the sample into a 150 ml beaker, add 1 ml of low level ISA, and 
place the beaker on the magnetic stirrer. Begin stirring. Lower the electrode tips into the 
solution. After the reading has stabilized, record the mV reading and determine the 
concentration from the low level calibration curve. 
  
10.  Prepare a new low level calibration curve daily. Check the calibration curve every two 
hours by repeating Steps 2-7. 
 
TITRATION 
 
Titration is a very accurate determination of total sulfide or total silver concentration. This method 
makes use of the electrode as an endpoint detector. The electrode can also be used to determine 
halide concentrations.