dhi molstic-s supply only platform 14 in. and 12 in. pns 401876 and 401877 installation sheet Manual Do Utilizador

Página de 2
 
 
 
molstic-S™ Supply Only Platform 
1/4 in. and 1/2 in. 
P/Ns 401876 and 401877 
Installation Sheet 
 
The 1/4 in. and 1/2 in. supply only molstic-S platforms, 
P/N 401876 and 401877, provide inlet gas pressure 
regulation to either the Device Under Test (DUT) or the 
downstream molstic-S mounting systems.  They are 
used to control the pressure supplied to an upstream 
DUT, or to adjust the flow rate through the system 
while calibrating a downstream DUT. 
 
 
High pressure gases are potentially hazardous.  
Energy stored in these gases can be released 
unexpectedly and with extreme force.  High 
pressure systems should be assembled and 
operated only by personnel who have been 
instructed in proper safety practices. 
 
 
 
Figure 1  1/4 in. Supply Only molstic-S 
1. Quick 
connect 
adaptor 
(includes stem with 1/4 in. SWG - not shown) 
2. Pressure 
reducing 
regulator 
3.  Flow metering/shut-off valve 
4.  Outlet adaptor, 1/4 in. tube x 1/4 in. male VCR 
 
 
Figure 2  1/2 in. Supply Only molstic-S 
1.  Quick connect adaptor 
(includes stem with 1/2 in. SWG - not shown) 
2. Pressure 
reducing 
regulator 
3.  Flow metering/shut-off valve 
4.  Outlet adaptor, 1/2 in. tube x 1/2 in. male VCR 
 
OPERATION OF SUPPLY ONLY MOLSTIC-S 
  Set regulator outlet pressure to minimum. 
 
Rotate the knob of the pressure reducing regulator 
in the counterclockwise direction until the spring 
force of the diaphragm is relieved.  Insure that the 
shut-off valve is in the closed position. 
 
 
Be sure that the molstic-S supply only 
platform shut-off valve is closed prior to 
connecting the quick connect stem to the 
quick connect body (located on the inlet side 
of the regulator). 
 
 
  Connect the gas supply. 
 
Connect the gas supply to the quick connect stem 
(either 1/4 in. or 1/2 in. tube).  The gas supply 
pressure should not exceed 1.7 MPa (250 psig).  
Insert the stem into the quick connect body. 
  Connect the outlet of the supply only platform to 
either the DUT or the inlet of the molstic-S.