Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Therapy Modes and Controls 
 
 
 
Control Modes 
19 
 
Control Modes 
A/C-PC: Assisted/Control-Pressure Control 
D
ESCRIPTION
:
 
 
The A/C-PC mode provides pressure-controlled mandatory or assist-control breaths. When the Trigger Type is set to Off, the ventilator triggers and 
cycles all breaths. When the Trigger Type is not set to Off, then the ventilator or the patient can trigger a breath, and the ventilator cycles all 
breaths.   
S
ETTINGS
:
 
 
Setting Name 
Description 
Pressure Control 
Inspiratory pressure above PEEP 
PEEP 
Positive end expiratory pressure 
Rise Time 
Time required for the ventilator to change from the expiratory pressure setting to the inspiratory 
pressure setting when the breath is triggered. 
Breath Rate 
Minimum rate of breaths per minute 
Inspiratory Time 
Length of the inspiratory phase 
Trigger Type 
• 
Auto-Trak (passive circuits only) 
• 
Sensitive Auto-Trak (passive circuits only) 
• 
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) 
• 
Off 
Flow Trigger 
Sensitivity 
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the 
patient’s inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. 
Flow Cycle 
Sensitivity 
This control is available when the trigger type is Flow Trigger. 
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the 
device cycles to expiration. 
FiO
2
  
(optional) 
Requires model with oxygen blender 
S
ETTABLE ALARMS
• 
Circuit disconnect 
• 
High tidal volume 
• 
Low tidal volume 
• 
High minute ventilation 
• 
Low minute ventilation 
• 
High respiratory rate 
• 
Low respiratory rate 
I
LLUSTRATION
 
 
1.  EPAP 
2.  Pressure control