Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
45 
 
 
High-priority System Alarms 
V
ENTILATOR 
I
NOPERATIVE
 
Priority 
High 
Why it occurs 
The system self-test indicates a failure or malfunction of a component that causes therapy to stop or 
not meet essential performance criteria. 
What to do 
Contact Philips Respironics Customer Service 
Device 
performance 
during active 
alarm 
Therapy is stopped and both audible and visual alarms are continuous. Depending on the systems 
impacted, you may or may not see a message on the screen. 
 
V
ENTILATOR 
S
ERVICE 
R
EQUIRED
 
Priority 
High 
Why it occurs 
This alarm occurs when the system self-test indicates an error that does not affect the ability of the 
ventilator to meet essential performance criteria. 
What to do 
Contact Philips Respironics Customer Service 
Device 
performance 
during active 
alarm 
The device continues to operate (possibly in a reduced capacity mode). If the problem is not corrected, 
the device will generate a reminder message until the issue is corrected. Additionally, if therapy is 
stopped, a reminder message will immediately appear when therapy is turned on again. 
 
O
BSTRUCTION 
 
Priority 
High 
Why it occurs 
The ventilator detects an obstruction in the patient’s inhalation path, exhalation path, or external flow 
sensor. The ventilator detects that the exhalation port is missing. 
What to do 
• 
Check the circuit for kinked or pinched tubing. 
• 
Check the bacteria filter or HME for a blockage or occlusion. 
• 
Ensure that the exhalation port is not occluded or missing. 
• 
Check the external flow sensor for a blockage or occlusion. 
Requirements 
Passive, dual limb, active PAP, or active flow circuit 
Device 
performance 
during active 
alarm 
The device will automatically open the active exhalation valve and continue to operate. 
Algorithm 
summary  
Any circuit, inhalation limb: an obstruction is detected when either of the following conditions exist: 
• 
The flow exiting the machine is less than 0.5 lpm for 5 seconds continuously 
• 
The flow exiting the machine during inspiration is less than 1 lpm for 5 seconds, 2 breaths, or 65 
seconds for very low breath rates. 
 
Any circuit: The expiratory port is missing and causes an obstruction alarm if the average flow is less 
than 1 lpm for 2 breaths or 60 seconds. 
 
Active flow or dual limb circuit: an obstruction is detected when the external flow sensor measures 
less than 0.5 lpm for 65 continuous seconds. 
 
H
IGH 
E
XPIRATORY 
P
RESSURE
 
Priority 
High 
Why it occurs 
During the expiratory phase, the delivered pressure exceeds the target patient pressure by 5 cmH
2
O or 
more.  
What to do 
• 
Check the circuit for kinked or pinched tubing. 
• 
Ensure that the leak device is not blocked or occluded. 
 
Note: This alarm condition may be due to pinched tubing or the patient having a fast breath rate.