Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
49 
 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the calculated minute ventilation is greater than the low 
minute ventilation alarm setting. 
volume under delivery 
Device will continue to operate. 
Alarm Settings 
Adult and Pediatric patient types: 0.2 to 30 LPM in increments of 0.1 LPM. 
Infant patient type: 0.2 to 10 LPM in increments of 0.1 LPM. 
 
L
OW 
R
ESPIRATORY 
R
ATE
 
Priority 
High 
Why it occurs 
The patient’s respiratory rate is less than or equal to the low respiratory rate alarm setting. 
Or, no breath has occurred for 15 seconds. 
What to do 
• 
Check the circuit for kinked or pinched tubing. 
• 
Check the circuit for a leak or disconnect. 
Note: A patient that is sleeping or medicated may breathe at a lower rate. 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the measured respiratory rate is greater than the low 
respiratory rate alarm setting. 
 
Device will continue to operate. 
Alarm Settings 
Off, 1 to 80 BPM in increments of 1 BPM. 
 
H
IGH 
I
NSPIRATORY 
P
RESSURE
 
Priority 
Variable 
Why it occurs 
The alarm will sound if the measured patient pressure exceeds the high inspiratory pressure alarm 
setting. 
 
Occurs in stages, escalating from an audible single beep for the first two occurrences to an audible 
medium priority on the third occurrence, and then to a high priority alarm if the condition continues. 
Occurrences must be consecutive. 
What to do 
• 
Check the patient status 
• 
Check the ventilator for the following potential causes for this alarm: 
o
 
Kinked, pinched, or blocked tubing 
o
 
Blocked leak device or blocked exhalation device 
o
 
Secretions in the HME 
Device 
performance 
during active 
alarm 
Inspiration is cycled to PEEP with this alarm 
Device will continue to operate. 
The alarm is automatically resolved when the peak inspiratory pressure is less than or equal to the 
high inspiratory pressure alarm setting. The device will cycle out of the breath. 
Alarm Settings 
10 to 90 cmH
2
O in increments of 1 cmH
2
O. 
Algorithm 
summary  
The High Inspiratory Pressure alarm shall generate a High Priority alarm on the 10
th
 consecutive breath 
or if 30 seconds has elapsed since a previous breath and the Peak Inspiratory pressure from the 
current breath is greater than or equal to the alarm setting 
 
Medium-priority System Alarms 
C
IRCUIT 
L
EAKAGE
 
Priority 
Medium 
Why it occurs 
In the Active PAP circuit, a leak in the active exhalation valve is detected 
What to do 
• 
Check the valve and both lines for kinked or pinched tubing. 
• 
Check the circuit and both lines for a leak or disconnect at both ends. 
• 
Check that the valve is not damaged and can properly seal and open. 
Requirements 
Active PAP circuit is selected 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when the circuit is reconnected or the valve is fixed. 
 
Device will continue to operate.