Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Alarms and System Messages 
 
 
 
Alarms and System Messages 
53 
 
H
IGH 
O
XYGEN 
I
NPUT 
P
RESSURE 
 
Priority 
Medium 
Why it occurs 
The measured oxygen inlet pressure is greater than or equal to 87psig. 
What to do 
Check the O
source. 
Requirements 
Additional oxygen 
Device 
performance 
during active 
alarm 
Device will continue to operate. 
 
Medium-priority Patient Alarms with Variable Settings 
H
IGH 
T
IDAL 
V
OLUME
 
Priority 
Medium 
Why it occurs 
Passive, active flow, or dual limb circuit types: 
The estimated exhaled tidal volume is greater than or equal to the High Vte alarm setting for a number 
of consecutive breaths. The number depends on the therapy mode as follows.  
Three consecutive breaths: A/C-PC, CPAP, PSV, S/T, SIMV-VC, and SIMV-PC.  
Six consecutive breaths: A/C-VC 
 
Active PAP circuit type: 
The delivered tidal volume is greater than or equal to the High Vti alarm setting for three consecutive 
breaths. 
 
What to do 
• 
Check the patient for pain or anxiety. 
• 
Check the circuit for kinked or pinched tubing. 
• 
Ensure that the active exhalation valve is attached. 
 
Note: For active flow and dual limb, low flow O
2
 or the use of a nebulizer may increase tidal volumes 
beyond the alarm setting. 
Device 
performance 
during active 
alarm 
The alarm is automatically resolved when a breath occurs in which the exhaled tidal volume does not 
reach the High Vte alarm setting. 
 
Device will continue to operate. 
Alarm Settings 
10 to 2000 ml in increments of 5 ml. 
 
L
OW 
T
IDAL 
V
OLUME
 
Priority 
Medium 
Why it occurs 
Passive, active flow, or dual limb circuit types: 
The estimated exhaled tidal volume is less than or equal to the low tidal volume alarm setting for a 
number of consecutive breaths. The number depends on the therapy mode as follows.  
Three consecutive breaths: A/C-PC, CPAP, PSV, S/T, SIMV-VC, and SIMV-PC.  
Six consecutive breaths: A/C-VC 
 
Active PAP circuit type: 
The delivered tidal volume is less than or equal to the low tidal volume setting. 
What to do 
• 
Check the circuit for kinked, pinched, or blocked tubing. 
• 
Ensure the leak device is not blocked, occluded, or disconnected. 
• 
Ensure the diaphragm in the active exhalation device is inserted correctly. 
• 
Check the mask fit or change the mask. 
Requirements 
Passive, active flow, or dual limb circuit types: 
The alarm is automatically resolved when a breath occurs in which the exhaled tidal volume exceeds 
the low tidal volume alarm setting. 
 
Device will continue to operate. 
 
Active PAP circuit type