Keysight Technologies U1583BClip-on ammeter adapter 0.5 -400 A 32 mm U1583B Manual Do Utilizador
Códigos do produto
U1583B
Agilent U1583B Current Clamp
Operating Instructions
The U1583B current clamp is a dual range 40 A and 400 A clamp-on AC current clamp. This U1583B current
clamp is designed for Agilent handheld digital multimeter (DMM) and Agilent handheld oscilloscope. The
BNC-to-banana plug is required to connect the current clamp with DMM. While for handheld oscilloscope,
it needs to use with BNC connector.
Assistance
For technical assistance, contact your nearest Agilent Sales Offi ce or visit the Agilent website at
www.agilent.com/fi nd/assist for further information.
Regulatory Markings
The CE mark is a registered trademark of the European Community.This CE mark
shows that the product complies with all the relevant European Legal Directives. If
it was accompanied by a year, it indicates the year the design was approved. This
ISM device complies with Canadian ICES-001.
Product contains restricted substance(s) above the maximum value, with 40 yr
Product contains restricted substance(s) above the maximum value, with 40 yr
Environmental Protection Use Period.
The CSA mark is a registered trademark of the Canadian Standards Association.
The C-tick mark is a registered trademark of the Spectrum Management Agency of
Australia. This signifi es compliance with the Australia EMC Framework regulations
under the terms of the Radio Communication Act of 1992.
This instrument complies with the WEEE Directive (2002/96/EC) marking
This instrument complies with the WEEE Directive (2002/96/EC) marking
requirement. This affi xed product label indicates that you must not discard this
electrical/electronic product in domestic household waste.
Safety Information
Please use the Agilent U1583A current clamp only as specifi ed in this manual. Otherwise, the protection
provided may be impaired. A WARNING identifi es conditions and actions that pose hazards to the user. A
CAUTION identifi es conditions and actions that may be damaging to the equipment under test. To avoid
possible electric shock, personal injury or damage to this instrument, ensure that you use the adapter
safely, and refer to the guidelines below.
AC - Alternating Current
Range button in release mode.
Range ~ 400 A, Output ~ 1 mV/A
Caution, risk of danger (Refer to the
user’s and service guide for details)
DC - Direct Current
400A MAX
Maximum allowable current
measurement is 400 A
Ground
CAT III 600V
Category III 600V over-voltage protection
Double Insulation
Range button in lock mode.
Range ~ 40 A, Output ~ 10 mV/A
Caution, risk of electric shock (Refer to
the user’s and service guide for detail)
To be applied around or removed from
un-insulated hazardous live conductors
- Do not use the adapter if it is damaged. Inspect the case before you use the adapter.
Look for cracks or missing plastic. Pay particular attention to the insulation
surrounding the connectors.
- Inspect the clamp jaw before each use. It shall not have cracks or missing parts,
loose or weakened components. Be sure there is insulation surrounding the jaw.