9 Instant help
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible
from power supply. The weighing process must then be restarted from the beginning.
Help:
Fault Possible
cause
The displayed weight does not glow.
• The balance is not switched on.
• The mains supply connection has been
interrupted (mains cable not plugged
in/faulty).
in/faulty).
• Power supply interrupted.
•
Batteries are inserted incorrectly or empty
• No batteries inserted.
The displayed weight is permanently
changing
changing
• Draught/air movement
• Table/floor vibrations
• The weighing plate is in contact with
foreign matter.
• Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
interfering device if possible)
The weighing result is obviously
incorrect
incorrect
• The display of the balance is not at zero
• Adjustment is no longer correct.
• Great fluctuations in temperature.
•
•
Electromagnetic fields / static charging
(choose different location/switch off
interfering device if possible)
interfering device if possible)
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.