Chapter 2
Configuration and Installation
© National Instruments Corporation
2-7
SC-2040 User Manual
To close a switch, push down the number side.  Closing switch A selects a gain of 500.  Closing
switch B selects a gain of 200.  Closing switch C selects a gain of 100.  Closing switch D selects
a gain of 10.  Opening all the switches selects a gain of 1.  You can select other gains by closing
more than one switch, as shown in Table 2-7.
Table 2-7.  Switch Settings for Gain Selection
Desired Gain
Close Switches
1
None
10
D
100
C
200
B
300
B and C
500
A
600
A and C
700
A and B
800
A, B, and C
Input Mode Selection
Jumpers W2 through W9 select the input mode for each channel on the SC-2040.  Position A-B
leaves the negative input of the instrumentation amplifier connected only to the front connector.
This is the factory-default setting.  Position B-C connects the negative input of the
instrumentation amplifier to the board analog ground through a 100 k
Ω
 resistor.  This setting is
useful for keeping floating, or nonground-referenced, sources from saturating the instrumentation
amplifier.  Table 2-8 shows the input mode jumper and channel selections.
Table 2-8.  Input Mode Selection Jumpers
Channel
Use Jumper
0
W2
1
W3
2
W4
3
W5
4
W6
5
W7
6
W8
7
W9