Agilent Technologies E2625A Manual Do Utilizador

Página de 48
2- 10
Getting Started and Using Masks
To align a Mask to the waveform
To align a Mask to the waveform
1
Access the Mask Test Setup dialog box and make sure the mask test is 
enabled.
2
Locate the 
Use File Setup When Aligning
 option.
When aligning a mask to the waveform, these settings are used:
• Channel range and offset
• Time base range and position
• Trigger mode and level
• Mask test scaling source, position(s), and delay
You can either use the current oscilloscope settings, or you can use the standard 
settings defined in the mask file.
3
If you want to use the standard settings defined in the mask file, 
checkmark the 
Use File Setup When Aligning
 option.
4
If you want to use the current oscilloscope settings, do not checkmark 
this box. 
5
Click the 
Align
 button to align the mask.
When aligning a 
mask, you can use 
the current settings 
for the oscilloscope 
setup, or use the 
standard settings in 
the mask file.