Agilent Technologies E2625A Manual Do Utilizador
2- 10
Getting Started and Using Masks
To align a Mask to the waveform
To align a Mask to the waveform
1
Access the Mask Test Setup dialog box and make sure the mask test is
enabled.
enabled.
2
Locate the
Use File Setup When Aligning
option.
When aligning a mask to the waveform, these settings are used:
• Channel range and offset
• Time base range and position
• Trigger mode and level
• Mask test scaling source, position(s), and delay
You can either use the current oscilloscope settings, or you can use the standard
settings defined in the mask file.
settings defined in the mask file.
3
If you want to use the standard settings defined in the mask file,
checkmark the
checkmark the
Use File Setup When Aligning
option.
4
If you want to use the current oscilloscope settings, do not checkmark
this box.
this box.
5
Click the
Align
button to align the mask.
When aligning a
mask, you can use
the current settings
for the oscilloscope
setup, or use the
standard settings in
the mask file.
mask, you can use
the current settings
for the oscilloscope
setup, or use the
standard settings in
the mask file.