Agilent Technologies E2625A Manual Do Utilizador

Página de 48
2- 14
Getting Started and Using Masks
To stop the Mask Test on the first failure
To stop the Mask Test on the first failure
1
Access the Mask Test Setup dialog box and make sure the mask test is 
enabled.
2
Click 
Stop On Failure
 to checkmark the box.
3
Run the mask test.
When the mask test encounters the first failure, the test will stop and the failure 
result will be displayed in the Mask Test tab description area.
Analyzing a failure
If your waveform crosses the boundary of the mask violation region, the 
waveform cross-over areas will be displayed in red.  All waveform locations 
displayed in red indicate that the waveform fails the mask test in these areas.  
If the waveform just “touches” the boundary, the mask test fails.
Standard masks and the designated violation regions are shown in chapter 3.
When you start the mask test with 
Stop On Failure enabled, if Infiniium 
encounters a failure, the oscilloscope acquisition system will be stopped.