Patient Safety Tests
Testing the Monitor  
4-39
Patient Safety Tests
Philips safety tests meet the standards of, and are performed in accordance with IEC 60601-1, 
Clause 19 (EN60601-1, Second Edition, 1988; Amendment 1, 1991-11, Amendment 2, 1995-
03). 
The C3 patient monitor is a Class I device. It requires a protective earth (ground) wire. Keep 
this in mind when performing the following test procedures.
There are two categories of safety tests:
Ground Integrity
Electrical Leakage
Ground 
Integrity
Table 4-3.  Ground Integrity
Test or “Inspection” to Perform
Expected Test Results
Protective Earth
With mains cable:
Maximum impedance = x <= 100 mOhms