14
VENTURE™
PART NO. 1093845
SPECIFICATIONS
OPERATING VOLTAGES
EXTERNAL AC/DC POWER
Output: 12 + 3 VDC, 100 mA
 ADAPTER:
Connector: 5.5mm OD x 2.1mm
ID x 12.0mm No. 2 Connector
ALARMS
Low Battery (Two [2] levels)
Cannula Disconnect (Breath not
sensed)
Device Malfunction
Low / No Gas Indicator
INDICATORS/SYMBOLS:
O
2     
  LED:
Pulse Delivery
LED:
Cannula  Disconnect/Missing
Breath Indicator, Audible Alarm
LED:
Battery Low Charge Indicator (1/2 to 2
hours of use time left), Audible Alarm
       
LED:
Low or No Gas, Audible Alarm
:
Class BF
BATTERY:
Alkaline: (1) "D" size cell
Approx. Battery Usage Time:
80 Hours @ 20 bpm
(Variable Breath Mode)
OXYGEN DOSE DURATION:
1-variable duration breathing:
Auto-adjusts between 0.5 and 1.0
seconds based on patient breathing rate.
3-fixed duration breath mode:
Fixed 1.0 seconds, 0.75 seconds,
0.50 seconds
FLOWRATE RANGE (L/Min.):
0 to 6 L/Min.
OXYGEN SOURCE PRESSURE:
Venture 20:
20 + 3 psig (138 + 20 kPa)
Venture 50:
50 + 5 psig (345 + 35 kPa)
REGULATOR
Flow Settings:
0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0,
5.0, 6.0  L/Min.
Regulated Pressure:
20 psig
O
2