Testing Millivoltage (mV) Function 
To verify accuracy of the mV function, do the following. 
1.   Connect the Calibrator to the   and COM inputs on the meter. 
2.   Turn the rotary switch to mV. 
3.   Set the calibrator for the voltage from step 1 to 3 in Table 3. 
4.   Compare the reading on the Meter display with the display reading in Table 3. 
      If the display reading falls outside of the range shown in Table 3, the Meter does not 
      meet specification. 
 
Table 3. DC mV Test 
Input Reading 
Step Voltage 
61-633 
61-635 
36.000mV 
35.848 to 36.156 
35.894 to 36.102 
-36.000mv 
-35.848 to -36.156 
-35.894 to -36.102 
360.00mV 
359.08 to 360.92 
359.58 to 360.42 
 
5.   Press the blue button on the Meter to toggle to AC mV function. 
6.   Set the calibrator for the voltage and frequency from step 1 & 2 in Table 4. 
7.   Compare the reading on the Meter display with the display reading in Table 4.  
      If the display reading falls outside of the range shown in Table 4, the Meter does 
      not meet specification. 
 
Table 4. AC mV Test 
Input Reading 
Step Voltage 
Frequency 
61-633 
61-635 
360.00mV 
50Hz 
355.18 to 364.82 
356.98 to 363.02 
360.00mV 
1KHz 
352.30 to 367.70 
355.90 to 364.10 
 
 
Testing the PEAK HOLD Function 
To verify operation of the PEAK HOLD feature, do the following. 
1.   Connect the Calibrator to the V
Ω and COM inputs on the Meter. 
2.   Apply 1.0v ac rms at 60Hz from the Calibrator to the V
Ω and COM inputs of the Meter. 
3.   Turn the rotary switch to V. 
Note:  The rms converter is not used in Peak Hold mode. The digital display represents the  
           actual peak value of the input. 
4.   Press (PEAK H) and then (M/M/A). 
5.   The reading on sub-display should be within 1.441 (Max) and -1.441 (MIN). 
6.   Press (PEAK H) to escape. 
 