Table 9. AC mA Test
 
Input Readings 
Step Current Frequency 
61-633 
61-635 
36.000mA 
50 Hz 
35.488 to 36.512 
35.632 to 36.368 
36.000mA 
400 Hz 
35.488 to 36.512 
35.632 to 36.368 
360.00mA 
50 Hz 
354.88 to 365.12 
356.32 to 363.68 
360.00mA 
400Hzz 
354.88 to 365.12 
356.32 to 363.68 
 
 
Testing the Amp (A) Function 
To verify the accuracy in the ampere (A) measurement function, do the 
Following: 
1.   Connect the Calibrator to the A and COM inputs of the Meter. 
2.   Turn the rotary switch to A 
3.   Apply the inputs for steps 1-2 in Table 10. 
4.   For each input, compare the readings on the Meter display to the display readings in  
      Table 10. 
5.   If the display reading falls outside of the range shown in Table 10, the meter does not  
      meet specification. 
 
Table 10. DCA Test 
Input Readings 
Step Current 61-633 
61-635 
3.6000A 
3.5780 to 3.6220 
3.5888 to 3.6112 
9.000A 
8.9150 to 9.0850 
8.9420 to 9.0580 
 
6.   Press the blue button on the Meter to toggle to the AC amp measurement function. 
7.   Apply the inputs for steps 1-4 in Table 11. 
8.   For each input, compare the reading on the Meter display to the display readings in  
      Table 11. 
9.   If the display reading falls outside of the range shown in Table 11, the meter does not 
      meet specification. 
 
Table 11. ACA Test
 
Input Readings 
Step Current  Frequency 
61-633 
61-635 
3.6000A 
50 Hz 
3.5488 to 3.6512 
3.5632 to 3.6368 
3.6000A 
400 Hz 
3.5488 to 3.6512 
3.5632 to 3.6368 
9.000A 
50 Hz 
8.8120 to 9.1888 
8.8480 to 9.1520 
9.000A 
400 Hz 
8.8120 to 9.1888 
8.8480 to 9.1520 
 
Testing the Frequency Function 
To verify the accuracy of the Meter’s frequency function, do the following: 
1.   Connect the Calibrator to the V
Ω and COM inputs on the Meter. 
Note:  The accuracy of the Calibrator’s frequency function must be appropriate for the specified accuracy of the      
Meter. 
2.   Set the rotary switch to Hz. 
3.   Set the Function Generator for the square wave voltage and frequency for steps 1-5 of Table 12. 
4.   Compare the reading on the Meter display with the display reading shown in Table 12. 
5    If the display reading falls outside of the range shown in Table 12, the Meter does not  
      meet specification. 
 
 
Form Number TM61633-5 
Rev 5 Dec 2006