Testing the Resistance Function 
To verify the accuracy of the resistance function, do the following: 
1.   Connect the Calibrator to V
Ω and COM on the Meter. 
2.   Turn the rotary switch to 
Ω 
3.   Apply the inputs for step 1-7 in Table 5. 
      *Set reference if using a Multifunction Calibrator with diff or zero mode for steps 1, 2, and 3. 
      Compare the Meter display readings to the display readings in Table 5 step 1-7. 
4.   If the display reading falls outside of the range shown in Table 5, the meter does not  
      meet specification. 
 
Table 5. 
Ω Test 
Input 
Readings 
Step 
Resistance 
Ω 
61-633 61-635 
0.00 
00.00 to 00.02 
00.00 to 00.02 
360.00 
358.00 to 362.00 
358.72 to 361.28 
3.6000K 
3.5800 to 3.6200 
3.5872 to 3.6128 
36.000K 
35.800 to 36.200 
35.872 to 36.128 
360.00K 
358.00 to 362.00 
358.72 to 361.28 
3.6000M 
3.5780 to 3.6220 
3.5852 to 3.6148 
36.000M 
34.150 to 37.850 
34.150 to 37.850 
 
5.   Press the blue button on the meter to toggle to LV
Ω function. 
6.   Apply the inputs for step 1-6 in Table 6. 
      *Set reference if using a Multifunction Calibrator with diff or zero mode for steps 1 & 2.  
      Compare the Meter display readings to the display readings in Table 6. 
7.   If the display reading falls outside of the reading shown in Table 6, the meter does not  
      meet specification. 
 
Table 6. LV 
Ω 
Input Readings 
Step 
Resistance 
Ω 
61-633 61-635 
0.00 
00.00 to 00.02 
00.00 to 00.02 
3.6000K 
3.5620 to 3.6380 
3.5764 to 3.6236 
36.000K 
35.620 to 36.380 
35.764 to 36.230 
360.00K 
356.20 to 363.80 
357.64 to 362.36 
3.6000M 
3.5600 to 3.6400 
3.5744 to 3.6256 
36.000M 
33.430 to 38.570 
33.430 to 38.570 
 
Testing the Capacitance Function 
The Meter measures capacitance by charging the capacitor with a known direct current, and measuring the 
resultant voltage, and calculating the capacitance. If the same capacitance is measured on an impedance bridge, a 
different reading may result. This variance is likely to be greater at higher frequencies. 
 
To verify the accuracy of the capacitance measuring function, do the following: 
1.   Apply the Capacitor to the V
Ω and COM inputs on the Meter. 
2.   For steps 1 through 8 in Table 7. 
      a)  Turn the rotary switch 
      b)  Set the V
Ω and COM inputs un-connected 
      c)  Select  
Δ or (REL) for steps 1 and 2 in Table 7. After steps 1 and 2 are complete, escape  
           from 
Δ mode and press (RANGE) key until (AUTO) is displayed. 
 
 
 
Form Number TM61633-5 
Rev 5 Dec 2006