Chapter 3
Wiring the Transmitter to the Sensor
3.1
Overview
This chapter describes how to connect Micro Motion Model 1700 and 2700 transmitters to a Micro 
Motion sensor. 
Note: If you have an integral installation, this step is not required. Continue with wiring the 
transmitter outputs (Chapters 4–
7).
Wiring requirements between the sensor and transmitter depend on the installation type (see 
Figure 2-1
)
.
If you have a 4-wire remote transmitter installation, review the information on 4-wire cable in 
Section 3.2, then follow the instructions in Section 3.3.
If you have a 9-wire remote transmitter installation, review the information on 9-wire cable in 
Section 3.2, then follow the instructions in Section 3.4.
If you have a remote core processor with remote transmitter installation, review the 
information on both 4-wire and 9-wire cable in Section 3.2, then follow the instructions in 
Section 3.5.
3.2
Cable types
This section describes the types of 4-wire cable and 9-wire cable that can be used for wiring the 
transmitter to the sensor.
CAUTION
Large electromagnetic fields can interfere with flowmeter communication 
signals.
Improper installation of cable or conduit can cause measurement error or flowmeter 
failure. To reduce the risk of measurement error or flowmeter failure, keep cable or 
conduit away from devices such as transformers, motors, and power lines which 
produce large electromagnetic fields.