Справочник Пользователя для A&D AD-4212A/B
30
Class Item
and
Parameter
Description
0
3-level comparator contact
output
output
Cpout
Comparator
output
Comparator
output
Cp-t
Comparator output
setting
Comparator output
setting
1
5-level comparator contact
output
output
Only the AD-4212B series
balance displays this class and
item.
balance displays this class and
item.
0
No comparison
1
Comparison, excluding “near zero” when stable or overloaded
2
Comparison, including “near zero” when stable or overloaded
3
Continuous comparison, excluding “near zero”
Cp
Comparator mode
Comparator mode
4
Continuous comparison, including “near zero”
0
Not added
Cp-r
*2
Comparison results
1
Added
Use in A&D standard format.
0
OFF
bep
LL buzzer
1
ON
Displayed only when 5-level
comparator output (
comparator output (
*3
) is selected.
0
OFF
bep_
LO buzzer
1
ON
0
OFF
bep-
OK buzzer
1
ON
0
OFF
bep
HI buzzer
1
ON
0
OFF
Cp fnc
Comparator
Comparator
bep
HH buzzer
1
ON
Displayed only when 5-level
comparator output (
comparator output (
*3
) is selected.
Cp HH
Secondary upper limit
Secondary upper limit
Cp Hi
Upper limit
Upper limit
Cp lo
Lower limit
Lower limit
Cp ll
Secondary lower limit
Secondary lower limit
Refer to "10-10 Comparator
Function”.
Function”.
Displays Cp HH,, Cp ll only when
5-level comparator output (
5-level comparator output (
*3
) is
selected.
0
Key mode (when stable)
Accepts the PRINT key or external
contact input to output (or store)
data only when the display is
stable.
contact input to output (or store)
data only when the display is
stable.
1
Auto print mode A
(Reference = zero)
(Reference = zero)
2
Auto print mode B
(Reference = last stable value)
(Reference = last stable value)
Outputs (or stores) data when the
display is stable and conditions of
ap-p, ap-b and the reference value
are met.
display is stable and conditions of
ap-p, ap-b and the reference value
are met.
3
Stream mode /
Interval memory mode
Interval memory mode
With data 0, outputs data
continuously; with data 2, uses
interval memory.
continuously; with data 2, uses
interval memory.
4
Key mode B (instantly)
Accepts the PRINT key or external
contact input to output (or store)
data, regardless of the display
condition.
contact input to output (or store)
data, regardless of the display
condition.
dout
Data output
Not displayed
when BCD
output (OP-01)
is installed.
prt
Data output mode
5
Key mode C (when stable)
Accepts the PRINT key or external
contact input to output (or store)
data, instantly when stable or after
the display is stabilized.
contact input to output (or store)
data, instantly when stable or after
the display is stabilized.
Factory setting
*2 This item is not displayed for the AD-4212B series balance or when BCD output (OP-01) is
installed.
*3 AD-4212A series balance: when OP-04 is installed
AD-4212B series balance: when “Cp-t” is set to “1”