Справочник Пользователя для Honeywell SC1000

Скачать
Страница из 174
 
SC Series Instruction Manual     page 135
DA
C Ou
tpu
t Cha
nnel
14
14.6  Analog Output Configuration
14.6.1 Identifying the 
Output Type
A DAC Output channel is available with one of two types of digital-to-analog
(DAC) outputs: voltage or current.  You can determine which type of output a
channel has by one of three ways:
Consulting the instrument’s Customer Information Sheet
Examining the 
SYSTEM MENU
 -> 
CONFIGURATION
 -> 
CHANNEL
 nn 
TYPE
 menu item where 
nn is the number of the channel. If the channel’s type is 
VOLTAGE DAC
, it has a 
voltage output. If the channel’s type is 
CURRENT DAC
, it has a current output.
Examining the channel’s circuit board as shown in the figure below.
14.6.2 Channel Menu Items
The Analog Output can be driven by any channel’s track, peak or valley value. 
See the “Channel Menu” section earlier in this chapter for a complete listing of
SETUP menu items available on the 
DAC. SETUP
 sub-menu. 
14.6.3 Output Selection
Jumpers located on the channel’s circuit board determine what outputs are gener-
ated when the value selected to drive the Analog Output (from the 
DAC. CHANNEL
 and
DAC. SOURCE
 menu items) equals the 
DAC. FULL SCALE
 and 
DAC. ZERO SCALE
 settings.
Figure 14-1: Digital-to-Analog Output Jumper Locations
DAC. ZERO SCALE
Output
DAC. FULL SCALE
Output
P31
jumper
J11
jumper
CHANNELS WITH VOLTAGE OUPUT
0-5V
2.5 Volts
5 Volts
bottom
open
0-5V
0 Volts
5 Volts
bottom
closed
±5V
0 Volts
5 Volts
top
open
0-10V
5 Volts
10 Volts
middle
open
0-10V
0 Volts
10 Volts
middle
closed
CHANNELS WITH CURRENT OUTPUT
4-20mA
4 mA
20 mA
bottom
closed
4-20mA
12 mA
20 mA
bottom
open