• Once the controller is calibrated to a solution by referring to its
value at 25 °C, all subsequent measurements are temperature
compensated to 25 °C. You can also obtain temperature compen-
sation to a different temperature reference point by calibrating
the meter to that value. For example, the conductivity and TDS
value of HI 7031L and HI 70442L at 20 °C are 1.28 mS/cm
and 1358 ppm, respectively. By adjusting the display to these
two values, all subsequent measurements are compensated to a
base temperature of 20 °C.
Calibration is now complete.
Make sure that the probe is properly installed and calibrated (see
the preceding pages).
Turn the switch to the "SET" position (see 8 - Functional Diagram).
The display will show the previously adjusted  (e.g. 1.60 mS/cm or
1400 ppm).
Using a small screwdriver, first adjust the setpoint through the
COARSE trimmer and then fine tune it to your required level with the
FINE trimmer (see 4 and 3 -  Functional Diagram) until the desired
value is displayed (e.g. 2.1 mS/cm or 1700 ppm).
ADJUSTEMENT OF SETPOINT
ADJUSTEMENT OF SETPOINT
ADJUSTEMENT OF SETPOINT
ADJUSTEMENT OF SETPOINT
ADJUSTEMENT OF SETPOINT
SET
READ
mS
EC
1.60
FINE
COARSE
SET POINT
mS
EC
SLOPE
CAL
ZERO
CAL
1.60
FINE
COARSE
SET POINT
mS
EC
SLOPE
CAL
ZERO
CAL
2.10