AMMONIA LOW RANGE
SPECIFICATIONS
Range
0.00 to 3.00 mg/L (as NH
3
-N)
Resolution
0.01 mg/L
Precision
±0.03 @ 1.50 mg/L
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter
@ 420 nm
Method
Adaptation of the 
ASTM Manual of Water and
Environmental Technology, D1426-92, Nessler
method. The reaction between ammonia and
reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code
Description
Q.ty/test
Q.ty/set
*
Reagent Vial
1 vial
25 vials
HI  93764-0
Nessler Reagent
4 drops
1 bottle
*
 Reagent Vial identification: A, white cap.
Note: Store the unused vials in their container in a cool and dark
place.
REAGENT SET
HI 93764A-25 Reagents for 25 tests
For other accessories see page 77.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Ammonia LR on the
secondary LCD by pressing PROGRAM t and s.
• Remove the cap from a Reagent Vial.
• Add exactly 5.0 mL of sample to the
vial, while keeping the vial at a 45-
degree angle.
• Replace the cap and mix by inverting
the vial a couples of times. This is
the blank.
Ammonia LR
59
PHOSPHORUS, ACID HYDROLYZABLE
SPECIFICATIONS
Range
0.00 to 5.00 mg/L (as PO
4
3-
)
Resolution
0.01 mg/L
Precision
±0.06 @ 2.50 mg/L
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter
@ 610 nm
Method
Adaptation of the EPA method 365.2 and 
Standard
Methods for the Examination of Water and
Wastewater, 20
th
 edition, 4500-P E, ascorbic acid
method. A mild acid digestion converts condensed
inorganic forms of  phosphates to orthophosphate.
Then the reaction between orthophosphate and
the reagents causes a blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Q.ty/test
Q.ty/set
*
Reagent Vial
1 vial
50 vials
---
NaOH Solution 1.20 N
2 mL
1 bottle
HI  93758-0
Phosphorus Reagent
1 packet
50 packets
*
 Reagent Vial identification: P, white cap.
Note: Store the unused vials in their container in a cool and dark place.
REAGENT SET
HI 93758B-50 Reagents for 50 tests
For other accessories see page 77.
REQUIRED ACCESSORIES
C 9800-01
Hanna reactor (115 VAC)
C 9800-02
Hanna reactor (230 VAC)
HI 740216
test tube cooling rack (25 holes)
HI 740217
Laboratory bench safety shield
For other accessories see page 77.
MEASUREMENT PROCEDURE
Before starting to use the reagent kit it is recommended
to read carefully all the instructions and the Health &
Safety Data Sheet (HSDS). Pay particular attention to all
warnings, cautions and notes. Failure to do so may result
in serious injury to the operator.
Phosphorus, Acid Hydrolyzable