Справочник Пользователя для Hanna Instruments c 214

Скачать
Страница из 40
25
AMMONIA HIGH RANGE
SPECIFICATIONS
Range
0 to 100 mg/L (as NH
3
-N)
Resolution
1 mg/L
Precision
±3 @ 50 mg/L
Typical EMC
±1 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter
@ 420 nm
Method
Adaptation of the 
ASTM Manual of Water and
Environmental Technology, D1426-92, Nessler
method. The reaction between ammonia and
reagents causes a yellow tint in the sample.
REQUIRED REAGENTS
Code
Description
Q.ty/test
Q.ty/set
*
Reagent Vial
1 vial
25 vials
HI  93764-0
Nessler Reagent
4 drops
1 bottle
*
 Reagent Vial identification: A, green cap.
Note: Store the unused vials in their container in a cool and dark
place.
REAGENT SET
HI 93764B-25
Reagents for 25 tests
For other accessories see page 77.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Ammonia HR on the
secondary LCD by pressing PROGRAM t and s.
• Remove the cap from a Reagent Vial.
• Add exactly 1.0 mL of sample to the
vial, while keeping the vial at a 45-
degree angle.
• Replace the cap and mix by inverting
the vial a couples of times. This is
the blank.
Ammonia HR
56
PHOSPHORUS, REACTIVE
SPECIFICATIONS
Range
0.00 to 5.00 mg/L (as PO
4
3-
)
Resolution
0.01 mg/L
Precision
±0.05 @ 2.50 mg/L
Typical EMC
±0.01 mg/L
Deviation
Light Source
Tungsten lamp with narrow band interference filter
@ 610 nm
Method
Adaptation of the EPA method 365.2 and 
Standard
Methods for the Examination of Water and
Wastewater, 20
th
 edition, 4500-P E, ascorbic acid
method. The reaction between orthophosphate and
the reagent causes a blue tint in the sample.
REQUIRED REAGENTS
Code
Description
Q.ty/test
Q.ty/set
*
Reagent Vial
1 vial
50 vials
HI  93758-0
Phosphorus Reagent
1 packet
50 packets
*
 Reagent Vial identification: P, red cap.
REAGENT SET
HI 93758A-50 Reagents for 50 tests
For other accessories see page 77.
MEASUREMENT PROCEDURE
• Select the program number corresponding to Reactive Phosphorus
on the secondary LCD by pressing PROGRAM t and s.
• Remove the cap from a Reagent Vial.
• Add exactly 5.0 mL of sample to the
vial, while keeping the vial at a 45-
degree angle.
• Replace the cap and mix by inverting
the vial a couple of times. This is
the blank.
Phosphorus, Reactive