Справочник Пользователя для eutech-instruments cyberscan con 11 conductivitytds meter

Скачать
Страница из 80
Instruction Manual 
 
CON 11/110 
 
11
4 CALIBRATION 
4.1 
Important Information on Meter Calibration 
Your meter has five measuring ranges. You have an option of calibrating your meter 
in a single point calibration for all the five ranges or for better accuracy; you can 
calibrate one point in each of the measuring ranges (up to five points). This selection 
of single point calibration and multi point calibration can be set in the SETUP menu 
Program P9.2. See page 52 . 
The following table lists the corresponding conductivity and TDS ranges. For either 
the single or multi point calibration, you should calibrate using a solution that falls 
between the values in the “recommended calibration solution range” column. 
Conductivity Range 
Recommended 
Calibration 
Solution Range 
 TDS 
Range 
Recommended 
Calibration Solution 
Range 
0 to 19.99 µS/cm, 
6.00 to 17.00 µS 
  0.00 to 9.99 ppm 
3.00 to 8.50 ppm 
20.0 to 199.9µS/cm,  
60.0 to 170.0 µS 
  10.0 to 99.9 ppm 
30.0 to 85.0 ppm 
200 to 1999 µS/cm; 
600 to 1700 µS 
  100 to 999 ppm; 
300 to 850 ppn 
2.00 to 19.99 mS/cm, 
6.00 to 17.00 mS 
  1.00 to 9.99 ppt 
3.00 to 8.50 ppt 
20.0 to 199.9 mS/cm
 
60.0 to 170.0 mS 
  10.0 to 99.9 ppt 
30.0 to 170.0 ppt 
Table 1: Recommended Calibration Solution Range in correspond to Conductivity / TDS  
 
In single point calibration, when you recalibrate your meter, the old calibration will be 
replaced by the new one even if the new calibration is done in a different range from 
the old calibration. For example, if you have previously calibrated your conductivity 
meter at 1413 µS in 0 to 2000 µS range, and you recalibrate at 12.88 mS in 0 to 
20.00 mS range, the new calibration will override the previous calibration in 0 to 2000 
uS range.    
In the case of multi point calibration, when you recalibrate your meter, old calibrations 
are replaced only on a range basis. For example, if you previously calibrated your 
conductivity meter at 1413 µS in 0 to 2000 µS range and you recalibrate at 1500 µS 
(also in 0 to 2000 µS range), the meter will replace only the old calibration data (1413 
µS) in that range.  The meter will retain all calibration data in other ranges.  
To completely recalibrate your meter, or when you use a replacement electrode, it is 
best to clear all calibration data. See page 45 for more information on calibration data 
clearance.