MAINTENANCE
CHANGING PROBE AND CALIBRATION
In case of probe is malfunctioned and need to be replaced with a new one, follow
below procedure to calibrate the probe to reach it’s fully performance.
below procedure to calibrate the probe to reach it’s fully performance.
To change the probe:
• Turn off the system.
• Screw loose the sealing nut, remove the malfunctioned probe from system.
• Connect the new probe , insure the indication point on both connector is
• Screw loose the sealing nut, remove the malfunctioned probe from system.
• Connect the new probe , insure the indication point on both connector is
matched, screw tight the sealing nut.
To calibrate:
• Turn off the system
• Press and hold both
• Press and hold both
and
simultaneously to enter factory mode.
• Select <Calibration>, press
.
-The calibration consists of two steps: bright scene calibration and dark
scene calibration.
scene calibration.
1. Brightscene:aiming the camera to a bright scene until whole screen is
pure white. Press
when ready. Holding still while calibrating. You
may lock the exposure value of a rather darker scene by press
,
this will help finding a bright scene.
2. Darkscene:aiming the camera to a dark scene until whole screen is
pure black. Press
when ready. Holding still while calibrating. You
may lock the exposure value of a rather bright scene by press
, this
will help finding a dark scene.
• When the calibration is done, restart the system.
24