Техническая Спецификация для Gw Instek LCR-821 Digital-Multimeter, DMM, LCR-821

Модели
LCR-821
Скачать
Страница из 91
  
APPENDIX 
131 
Angle (θ) Formula 
 
Background 
The Angle (θ) measures the phase on which 
Impedance (Z), Admittance (Y), Quality factor (Q), 
and Dissipation factor (D) are measured. 
Type 
Angle (θ) 
Formula 
jX
R
Z
S
 
C
j
R
L
j
R
 
D
Q
1
90
tan
1
 
cos
Z
R
S
 
sin
Z
X
S
 
jB
G
Y
P
 
L
j
G
C
j
G
 
Q
D
1
90
tan
 
cos
Y
G
P
 
sin
Y
B
P
 
 
   LCR-800 
User 
Manual 
132
 
Accuracy Definitions 
 
Primary Measurement Readout Error Formula 
2 counts±0.03%+0.02%[(1+Ka)
#
 or (X/Ymax )
#
 or 
(Ymin/X)
#
 ] (1+│D│)(1+Kb+Kc) 
2 counts±0.03%+0.02%[(1+Ka)
#
 or (X/Ymax )
#
 or 
(Ymin/X)
#
 ] (1+│Q│)(1+Kb+Kc) 
2 counts±0.03%+0.02%[(1+Ka)
#
or(X/Ymax)
#
or(Ymin/X)
#
 
] (1+ 1 /│Q│)(1+Kb+Kc) 
|Z| 
Ze = Depends on whether the component is a 
capacitor(C), resistor(R) or inductor(L): 
 
Circuit 
Formula for relevant circuit 
 
Ze = 2 counts±
0.03%+0.02%[(1+Ka)
#
or(X/Ymax)
#
or(Ymin/X)
#
 ] 
(1+│D│)(1+Kb+Kc) 
 R 
Ze = 2 counts±
0.03%+0.02%[(1+Ka)
#
or(X/Ymax)
#
or(Ymin/X)
#
 ] 
(1+│Q│)(1+Kb+Kc) 
 L 
Ze=2 counts±
0.03%+0.02%[(1+Ka)
#
or(X/Ymax)
#
or(Ymin/X)
#
]
(1+ 1 /│Q│)(1+Kb+Kc) 
 
Secondary Measurement Readout Error Formula 
D(C/D)  
2counts±0.0003 + 0.0002[(1+Ka)
#
 or (X/Ymax) 
#
 or 
(Ymin/X) 
#
] (1+│D│+D²)(1+Kb+Kc) 
Q(R/Q)  
2counts±0.0003 + 0.0002[(1+Ka) 
#
 or (X/Ymax) 
#
 or 
(Ymin/X) 
#
] (1+│Q│+Q²)(1+Kb+Kc) 
Q(L/Q) 
2counts±0.0003 + 0.0002[(1+Ka) 
#
 or (X/Ymax) 
#
 or 
(Ymin/X) 
#
] (1+│Q│+Q²)(1+Kb+Kc) 
θ(
 Z/
θ)  
θe=(180/π) x (Ze/100) 
R(C/R)  
D≧1 
2counts + 0.02%[(1+Ka)* or (Rx/Rmax)* or 
(Rmin/Rx)*] (1+ 1/│D│)(1+Kb+Kc)+0.03% 
 
D≦1 
2counts + 0.02%[(1+Ka)** or (Cx/Cmax)** or 
(Cmin/Cx)**] (1+ 1/│D│)(1+Kb+Kc)+0.03%