Справочник Пользователя для ITT SSV

Скачать
Страница из 56
13
  9. Install a new seal housing o-ring on the seal hous-
ing. A lubricant can be used to hold the o-ring on 
the diameter of the seal housing.
Carefully reinstall the seal housing between the 
pump and motor shaft and slide the seal housing 
down the shaft into position. Use caution when 
mounting the sealhousing between the  
pump/motor shaft so that the seal face  
on the stationary seat is not damaged.  
See step 9 in Figure 5.  
10. Reinstall the 4 hex cap screws, tightening  
the screws evenly and then torquing them  
to the values given in the engineering  
data. See step 10 in Figure 5.
11. Reinstall the coupling pin and locate the  
space shim on the shaft on the seal  
housing. If the shim is not available, a  
5 mm spacer can be used. See step 11 in  
Figure 5.  
12. Reinstall the coupling halves and evenly  
tighten the coupling bolts and torque  
them to the values given in the coupling  
section. Remove the spacer shim and save  
for future use. See step 1 in Figure 5.
13. Rotate the shaft by hand to insure that  
the pump and motor rotate freely.  
Reinstall the coupling guard. 
14. The pump and system should be vented  
prior to starting the pump. See Section 3,  
page 8, for venting procedures.
  
  
  1. Complete steps 1-4 as defined above for  
conventional seals.
  2. Loosen the 4 set screws located around the ID of 
the pump shaft.
  3. Remove the cartridge seal using the two tapped 
holes provided on the gland of the cartridge seal 
by threading two of the hex cap screws into these 
holes and evenly tightening these screws. Lift and 
remove the cartridge seal between the pump and 
motor shaft.   
  4. Inspect the pump shaft for any burrs or debris. 
Any burrs should be ground smooth with (fine 
grit) emery paper.
  5. Install a new o-ring on the turned fit of the new 
cartridge seal. Lubricate the o-ring located in the 
ID of the cartridge seal. Use a lubricant compat-
ible with the seal elastomers to hold the o-ring.
  6. Install the new cartridge seal on the pump by care-
fully sliding it between the pump and motor shaft 
and then sliding it into position.
  7. Complete steps 11-14 as defined above for the 
conventional seals.      
  8. After installing the coupling, tighten the four set 
screws located in the collar of the cartridge seal to 
secure the seal to the pump shaft.
  9. Rotate the shaft by hand to insure that the pump 
and motor rotate freely. Reinstall the coupling 
guard. 
10. Use vent connections on cartridge seal for proper 
venting. 
To remove the motor follow steps 1 through 4, as 
provided in the “MECHANICAL SEAL REPLACE-
MENT” section of this manual.
For motor frames 13TC and larger, remove the 4 
motor hex cap screws and the motor adapter. 
Install the motor adapter flange onto the new motor, 
torquing the 4 hex cap screws to the values provided 
in the “ENGINEERING DATA” section of this 
manual.
Complete the reassembly following steps 17 through 
 in the “MECHANICAL SEAL REPLACE-
MENT” section of this manual.
All additional unit service or maintenance, not ad-
dressed in this manual, should be performed at a 
qualified service location. Contact your local dealer 
or G&L Pumps distributor for assistance.
DISCONNECT AND LOCKOUT 
ELECTRICAL POWER BEFORE 
ATTEMPTING ANY MAINTE-
NANCE. FAILURE TO DO SO 
CAN CAUSE SHOCK, BURNS  
OR DEATH.
SYMPTOM
  MOTOR NOT RUNNING
 
See Probable Cause – 1 through 5
  LITTLE OR NO LIQUID DELIVERED BY PUMP
 
See Probable Cause – 6 through 1
  POWER CONSUMPTION TOO HIGH
 
See Probable Cause – 3, 1, 13, 15
  ExCESSIVE NOISE AND VIBRATION
 
See Probable Cause – 3, 6 - 8, 10, 1, 13, 16
PROBABLE CAUSE
  1. Motor thermal protector tripped.
  2. Open circuit breaker or blown fuse.
  3. Impellers binding.
  4. Motor improperly wired.
  5. Defective motor.
  6. Pump is not primed, air or gases in liquid.
  7. Discharge, suction plugged or valve closed.
  8. Incorrect rotation (three phase only).
  9. Low voltage or phase loss.
10. Impellers worn or plugged.
WARNING
Hazardous
voltage