24                                                                                                                     MT2000-0200-01 Rev f (10-2007) 
DCN0160
 
END (CFG MENU) 
RNG - Range 
LTP - Low Trim Point 
HTP - High Trim Point 
LTC -  Low Trim counts 
HTC - High Trim counts 
TMP - Module Temperature 
SOR - Signal Out of Range  
 
LRM - Low Range Margin 
 
LRS - Low Range mA 
SAT 
END 
LL1 - Liquid Level 1 
(Main Menu) 
Select 
UP/DOWN 
Arrow 
Configuration 2 (Hidden) Menu 
The sensor trim is now complete and the LRV has been set to equal 0 and URV has been set to the value of HTP. If 
other LRV and URV are desired, then proceed to the CAL menu and change LRV and URV accordingly. 
 
If the level cannot be moved, the transmitter must be removed for re-trimming.  Please contact the factory before 
trimming to confirm the procedure to follow. Should the trim be suspected of being corrupted, please consult the fac-
tory.  The transmitter can be brought back to the factory calibration conditions by performing a “3 Button Reset”.  
This is accomplished by simultaneously pressing “UP”, “DOWN” and “SELECT”.  This action is generally recom-
mended as a “last resort” prior to returning the transmitter to the factory for analysis, repair (if needed) and re-
calibration. 
 
Figure 9.2 
SAT 
ALRM 
9.2 
APPENDIX B: Additional Special Functions in the CFG2 (Hidden) Menu 
 
These functions are accessed by pressing the ‘UP’ and ‘DOWN’ buttons together, when the LCD display is on 
‘END’ at the end of the CFG menu.  Enter the Sensor Trim menu and scroll down to access the following func-
tions: