Technical data
Proline Promass 80 PROFIBUS PA
108
Pressure loss coefficient for Promass S, P
a0004607
Fig. 48: 
Pressure loss diagram for water 
Pressure loss (US units)
Pressure loss is dependent on fluid properties nominal diameter. Consult FMC Technologies for 
Applicator PC software to determine pressure loss in US units. All important instrument data is 
contained in the Applicator software program in order to optimize the design of measuring system. 
The software is used for following calculations:
•  Nominal diameter of the sensor with fluid characteristics such as viscosity, density, etc.
•  Pressure loss downstream of the measuring point.
•  Converting mass flow to volume flow, etc.
•  Simultaneous display of various meter size.
•  Determining measuring ranges.
The Applicator runs on any IBM compatible PC with windows.
DN
d[m]
K
K1
K3
8
8.31 
⋅ 10
–3
8.78 
⋅ 10
6
3.53 
⋅10
7
1.30 
⋅ 10
6
15
12.00 
⋅ 10
–3
1.81 
⋅ 10
6
9.99 
⋅ 10
6
1.87 
⋅ 10
5
25
17.60 
⋅ 10
–3
3.67 
⋅ 10
5
2.76 
⋅ 10
6
4.99 
⋅ 10
4
40
26.00 
⋅ 10
–3
8.00 
⋅ 10
4
7.96 
⋅ 10
5
1.09 
⋅ 10
4
50
40.50 
⋅ 10
–3
1.41 
⋅ 10
4
1.85 
⋅ 10
5
1.20 
⋅ 10
3
Pressure loss data includes interface between measuring tube and piping
1000
100
10
1
0.1
0.01
0.1
1
10
100
[mbar]
[t/h]
DN 40
DN 50
DN 25
DN 15
DN 8