Справочник Пользователя для FMC Corporation - Talstar 80

Скачать
Страница из 122
Commissioning
Proline Promass 80 PROFIBUS PA
60
6.6.4
Configuration examples with Simatic S7 HW-Konfig
Example 1:
a0004775
Fig. 33: 
Complete configuration using the Promass 80 GSD file. 
It is essential to adhere to the following sequence when configuring the modules in the 
PROFIBUS master (Class 1):
Slot 
sequence
Module
Byte length 
input data
Byte length 
output data
Description
1
AI
5
Analog Input function block 1
Output variable 
→ mass flow (factory setting)
2
AI
5
Analog Input function block 2
Output variable 
→ volume flow (factory setting)
3
AI
5
Analog Input function block 3
Output variable  
→ corrected volume flow (factory setting)
4
AI
5
Analog Input function block 4
Output variable  
→ density (factory setting)
5
SETTOT_MODETOT_TOTAL
5
2
Totalizer function block 1
TOTAL 
→ output variable = totalized mass flow (factory setting)
SETTOT 
→ totalizer control
MODETOT 
→ totalizer configuration 
6
DISPLAY_VALUE
5
Default value for local display
7
CONTROL_BLOCK
1
Control of device functions