Proline Promass 80 PROFIBUS PA
Troubleshooting
77
9.3
Process error messages
!
Note! 
See the information on 
9.3.1
Displaying the device status on PROFIBUS PA
Further information 
9.3.2
List of process error messages  
No.
Device status message
(local display)
PROFIBUS measured value status
Extended 
diagnostic 
message in the 
PROFIBUS 
master
Cause/remedy
Q
u
alit
y co
de 
(H
EX
)
Measured v
a
lue 
status
Qu
ality
 sta
tu
s
Qua
li
ty su
bs
ta
tu
s
Lim
its
P = Process error
$ = Fault message (with an effect on the outputs)
! = Notice message (without any effect on the outputs)
700
P: EMPTY PIPE
!: # 700
0x53
UNCERT
AIN
Sensor 
conversion 
not accurate
(measured 
value from 
sensor not 
accurate)
Constant
Empty pipe 
detected
Cause:
The process fluid density is outside the upper or lower 
limit values set for empty pipe detection.
Causes:
Air in the measuring tube Partly filled measuring tube 
Remedy:
1.
Ensure that there is no gas content in the process 
liquid.
2.
Adjust the upper or lower limit value for empty 
pipe detection to the current process conditions.
Access:
BASIC FUNCTIONS 
→ PROCESS PARAMETER → 
EPD PARAMETER 
→ EPD VALUE LOW or EPD 
VALUE HIGH
701
P: EXC. CURR. LIM
!: # 701
0x40
0x41
0x42
UNCERT
AIN
Non specific
(uncertain 
status)
O.K.
Low
High
Excitation too high
Cause:
The maximum current value for the sensor coils has 
been reached, since certain process fluid characteristics 
are extreme, e.g. high gas or solid content.
The instrument continues to work correctly.
Remedy:
In particular with outgassing fluids and/or increased 
gas content, the following measures are recommended 
to increase system pressure:
– Install the instrument at the outlet side of a pump.
– Install the instrument at the lowest point of an 
ascending pipeline.
– Install a flow restriction, e.g. reducer or orifice plate, 
downstream from the instrument.